Standby mode:
Standby mode is activated after approx. 10 minutes with no
activity on the station (input via touchscreen or opening/
closing of a module cover).
When the station switches over to standby mode, any user
logged in will be logged off automatically. The user will
need to log back in again when switching back to operating
mode.
The touchscreen is switched off in standby mode.
The charging function of X-am 125+ modules with charging
function is not affected by standby mode. The charging
process is not interrupted.
To switch to operating mode:
Tap the function key or
touch the touchscreen or
open or close a module flap.
4.2

Initial setup of station

1. Switch on the station, see chapter 4.1 on page 20.
2. Log in with the preconfigured "admin" user
(user name: admin, password: 123456), see chapter 4.5
on page 21.
3. Configure the test gas inlet, see chapter 4.6 on page 22.
4. Change language if necessary:
a. Select
> System configuration > Language.
b. Select desired language.
c. Confirm selection with OK.
5. Set date and time where necessary:
a. Select
> System configuration > Date & time.
b. Make the desired settings.
c. Confirm settings with OK.
4.3

Touchscreen display

The buttons on the touchscreen display change dynamically
depending on the task being executed. To execute an action,
select the corresponding icon on the display.
Press the
key on the master at any time to access the start
screen.
4.4

Start and test screens

The keys on the start and test screen change dynamically
depending on the log-in status, single-mode status and the
number of modules in use. For more information see the
X-dock 5300/6300/6600 Technical Manual.
4.4.1
Symbols
Menu
Confirm
Dräger X-dock 5300/6300/6600
Select this button to access the
menu.
Select this button to confirm an
input or function.
Cancel
Back
Log user in or out
4.5

Logging users in or out

NOTICE
i
i
A User-ID is required to log in. This must be created
beforehand by the administrator.
A user with administrator rights is created by default:
User name: admin
Password: 123456
NOTICE
i
i
Dräger recommends changing the admin password
after initial start-up.
To log in a user:
1. Select
.
a. Select
.
b. Select the desired user name from the list.
or
a. Select Select user.
b. Enter desired user name.
2. Enter the password and confirm with
NOTICE
i
i
When entering the user name, 3 already saved user
names will automatically be suggested and displayed.
For quick selection, please select the desired user
name.
To log out the current user:
1. Select
.
Information about the current user will be displayed.
2. Select
.
The current user will be logged out.
Basics
Select this button to cancel an
input or function.
Select this button to access the
previous screen.
Select this button to log users in
or out. The number in the
symbol indicates the permission
level (see chapter 6 on
page 26).
.
21