6.3
Charging function for gas monitors of the
X-am 125 series (optional)
Gas monitors of the X-am 125 series can be charged using the
X-am 125+ module. The station offers 2 different charging
functions:
Charge after 15 minutes of inactivity
Direct charging of a switched off gas monitor
The charging time is approx. 4 hours for a completely empty
battery. A new NiMH supply unit reaches its full capacity after
3 full charge/discharge cycles.
NOTICE
i
i
Never store the gas monitors for extended periods
(max. 2 months) without a power supply as the internal
buffer battery will run down.
To charge a gas monitor using a X-am 125+ module after a
test:
1. Place the gas monitor in the X-am 125+ module.
2. Close the module flap.
The instrument will be detected automatically.
After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status
LED.
The charging function starts automatically approx.
15 minutes after the last test.
The station additionally offers an option for directly charging
switched-off gas monitors with no waiting period. If this option
is activated, the gas monitors will not be automatically
switched on when inserted into the modules. The X-am 125+
modules then start charging directly.
To charge a switched-off gas monitor directly using an
X-am 125+ module:
1. Activate the option Do not switch on device on the station
(see Technical Manual).
2. Insert the switched-off gas monitor in the X-am 125+
module.
3. Close the module flap.
The gas monitor will be detected automatically and
charged directly.
If an error occurs:
Remove from the module and re-insert.
If this does not correct the error, have the module repaired.
CAUTION
!
A short-circuit in the charging contacts in the modules,
e. g. due to metal objects falling into the device, will not
damage the station but should be avoided due to a
potential risk of overheating and error displays on the
module.
Dräger X-dock 5300/6300/6600
Overview of charge status LEDs
Colour
Status
green
on continuously
green
flashing
red
flashing
6.4
Perform a firmware update
CAUTION
!
The station power supply must not be disconnected
during the installation process. The station may be
damaged if this is not observed.
NOTICE
i
i
The station does not support any USB data storage
device with an NTFS file system.
1. Download the firmware update from the internet:
a. Go to www.draeger.com.
b. Go to the X-dock product page and unzip the firmware
update to the root directory of an empty USB data
storage device.
CAUTION
!
The USB data storage device must not contain older
firmware files!
2. Connect USB data storage device with the firmware update
to the USB port of the station.
The USB icon is displayed in the status bar.
3. Select
>
configuration > Update.
A list of all firmware updates
available on the USB storage
device will be displayed.
4. Select
the
desired
update from the list. The selected
firmware update is marked blue.
Maintenance
Meaning
Charge status 100 %
Battery charging.
Charging error
System
firmware
27