9.4. Programming

The program P0 can be programmed. This position is ideal toe store the most frequent used parameters.
When the unit start-up these parameters of program P0 will be displayed. The program numbers of the
therapy suggestions P1 – P9 can be overwritten the same way.
Switch on
• Connect the standard main adapter to connector [1]
• Connect the main adapter plug to the mains socket
• Switch the unit on with the button [2]
Program selection
• Push the button [8]
• Start Selecting the program you wish to change P0 - P9
with one of the buttons [7] within 2½ seconds
• After finishing selections with the buttons [7] the
parameters will show after approx. 2 seconds
Treatment parameters: Adjusting or changing
• Adjust the parameters as decrypted under 9.1.
Set up the Sonopuls by hand" (follow the steps until
adjusting the intensity)
Storing
• Push at the program button [8] during minimal 3 seconds
• The parameter will be stored under the selected pro-
gram number. The storing will be confirmed by a
double beep

9.5. Contact control; sound signal on/of

Inadequate acoustic contact can be recognised by a sound signal; a short beep every 2½ seconds. As
desired this sound signal can be turned on or off.
Sound signal on/off
• Switch the unit of with the On/Off button
• Switch the unit ON while holding down the "duty cy-
cle button" [10] at the same time
• Release the button [10] as soon as in the upper dis-
play a "half 8" appears
In this way the sound signal is alternately turned on or
off *(default setting is "on").
23