ERROR INdICATORS ANd TROUBLESHOOTING TIPS

eRROR IndICAtORs
SyMBOL
Air plug disconnected .
Arm cuff not applied correctly .
Air is leaking from the arm cuff .
Cuff over or under inflated .
The arm cuff was inflated above 299
mmHg when inflating the cuff manually .
Movement during measurement .
Clothing is interfering with the
arm cuff .
Device error .
Cuff is not applied correctly .
Batteries are worn .
CAUSE
39
CORRECTION
Insert the plug securely . Refer to "Applying the
Arm Cuff" .
Apply the arm cuff correctly . Refer to "Applying
the Arm Cuff" .
Replace the cuff with the new one .
Repeat measurement . Remain still and do not
talk during measurement . Refer to "Taking a
Measurement" .
If "E2" appears repeatedly, inflate the cuff manually
until it is 30 to 40 mmHg above your previous
measurement result . Refer to "Taking a Measurement" .
Do not inflate the cuff above 299 mmHg . Refer to
"Taking a Measurement" .
Repeat measurement . Remain still and do not
talk during measurement . Refer to "Taking a
Measurement" .
Remove any clothing interfering with the arm cuff .
Refer to "Applying the Arm Cuff" .
Contact Customer Service .
Refer to "Applying the Arm Cuff" .
Replace the four batteries .
Refer to "Battery Installation" .