Avoid operating the product in the vicinity of sources of strong magnetic and electrical inter
ference (e.g. high-voltage lines, transmitters, transformer stations, computer tomographs,
magnetic resonance tomographs, etc.).
►
Use the product only in orthopaedic workshops, or during in-patient stays in rehabilitation
facilities.
NOTICE
Insufficient distance to HF communication devices (e.g. mobile phones, Bluetooth
devices, WiFi devices)
Product malfunction due to interference during data transfer.
►
Therefore, keeping the following minimum distances to these HF communication devices is
recommended:
• Mobile phone GSM 850 / GSM 900: 0.99 m
• Mobile phone GSM 1800 / GSM 1900 / UMTS: 0.7 m
• DECT cordless phones incl. base station: 0.35 m
• WiFi (routers, access points ...): 0.22 m
• Bluetooth devices (third-party products not approved by Ottobock): 0.22 m
NOTICE
Operating the product very close to other electronic devices
Limited function due to product malfunction.
►
Do not stack the product during operation nor operate it in the immediate vicinity of other
electronic devices in operation.
►
If simultaneous operation cannot be avoided, monitor the product and verify proper use in
the existing setup.
NOTICE
Use of unapproved accessories
>
Limited function due to product malfunction as a result of reduced interference resistance.
>
Interference of other electronic devices due to increased emissions.
►
Use the product only in combination with the accessories, signal converters and cables lis
ted in the section "Combination possibilities" (see Page 16).
INFORMATION
The BionicLink must be removed from the patient's prosthesis after adjustment.
INFORMATION
The BionicLink cannot recognise any defects on the prosthesis and prosthetic components by
itself. For this reason, attentive observation by the orthopaedic technician and knowledge about
the function of the prosthetic components are indispensable.
INFORMATION
The BionicLink is powered by the battery in the prosthesis. For this reason, adjustments should
only be made when the battery in the prosthesis is fully charged.
18 | Ottobock