Medela Thopaz Instrucciones De Uso página 37

Ocultar thumbs Ver también para Thopaz:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 325
Canister change
acoustic alarm
Visible on display
Error Problem
description
401
Battery low
306
Canister full
405
Standby mode
404
Leakage
402
USB connec-
tion not permit-
ted
406
Temperature
too high
305
Battery empty
302
System clogged – Check that tubing is
304
308
301
Leak in system
311
Selftest failed
314
313
Filter clogged
315
Thopaz over-
heated
Internal error
XXX
Operation
Troubleshooting instruc-
tions
Charge battery – connect
Thopaz to mains
Replace canister
Switch Thopaz on or off
– Check patient status
– Check system for leaks
– Continue with 'OK'
Unplug USB cable
Prepare backup device
Charge battery – connect
Thopaz to mains
clear/not twisted
– Replace canister
Check system for leaks
Snap the canister out and
in again
– Replace canister
– Continue with 'on'
Replace Thopaz
1. Switch Thopaz off and
on again
2. Inform Medela
Customer Service
Remarks / Potential source
of error
Remaining time of battery
approx. 30 min.
If the canister is not full,
see next page
In standby mode after
5 minutes
– Irregularity in the course
of therapy
– Indication of leakage
– See Chapter 6
USB connection not permit-
ted during operation or in
the vicinity of the patient
Do not expose Thopaz to
heat sources
– Tubing is kinked or
clogged, or filter in canis-
ter is clogged for approx.
10 min.
– Thopaz detects clotting of
the Thopaz tubing, NOT of
the patient catheter
– Disconnection
– Missing seal between
Thopaz and canister
Do not switch Thopaz on
if the patient is already
connected
Filter in canister clogged
If error message repeats,
inform Medela Customer
Service
Troubleshooting and alarms
Pressure
Yes
Yes
No
Yes
Yes
Yes
No
No
No
No
No
No
| 37

Publicidad

Tabla de contenido
loading

Tabla de contenido