Instructions For Fpt Test - Medela Thopaz Instrucciones De Uso

Ocultar thumbs Ver también para Thopaz:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 325

Instructions for FPT test

40 |
Warranty, maintenance and checks
Important: Patient must not be connected! Thopaz must be
disconnected from the mains (battery operation) and turned off.
1. Connect the tubing (079.0021) and the 0.8l (079.0016)
canister to Thopaz according to chapter 8.
2. Connect vacuum gauge with tube to the patient connector.
3.
Press simultaneously
(Tube e.g. ø 12 x 7 mm, not inclu-
ded in the scope of delivery)
3.1 Press and hold [
simultaneously, then...
3.2 Press additionally [
start safety test.
3.3 Follow the instructions
on the display.
0.8 l
], [
]
] to

Publicidad

Tabla de contenido
loading

Tabla de contenido