EMS SWISS DOLORCLAST smart Instrucciones De Empleo página 62

Tabla de contenido

Publicidad

TROUBLESHOOTING
The DolorClast smart provides an error code system in case of malfunction or incorrect handling:
1 LED blinking = Handpiece error; 2 LEDs blinking = Air pressure error; 3 LEDs blinking = Software error
Symptom
The display does not light up when the
unit is switched on
The display is blinking when the unit is
switched on
Handpiece error
Air pressure error
Pressure not adjustable
Pressure reduction during operation
Software error
50
Error code
Solutions
--
‡ Check the different electrical connections
‡ Check the mains electricity supply
‡ Check/replace the fuses and replace them with the type indicated on the back of the unit
‡ Send the unit to your authorized EMS repair center
--
‡ Send the unit to your authorized EMS repair center
1 LED
‡ Check the different connections of the unit
‡ Check the connection of the handpiece
‡ Test the unit with another handpiece
2 LEDs
‡ Check the different air connections
‡ Check the air pressure supply is within the values indicated on the back of the unit
‡ Check the compressor
‡ Check the connection between the handpiece and unit
‡ Test the unit with another handpiece
‡ Send the unit to your authorized EMS repair center
--
‡ Switch OFF/ON the unit
‡ Check the different air connections
‡ Check the air pressure supply is within the values indicated on the back of the unit
‡ Check the compressor
‡ Send the unit to your authorized EMS repair center
Repetitive
‡ Check the different air connections
"beep"
‡ Check the compressed air supply
‡ Contact your authorized EMS repair center
3 LEDs
‡ Send the unit to your authorized EMS repair center

Publicidad

Tabla de contenido
loading

Tabla de contenido