Lamira nebuliser handset
Nebuliser type
Weight of the nebuliser handset
Dimensions of the nebuliser handset (W x H x D)
Min. / Max. fill volume
a)
MMAD
(Mass Median Aerodynamic Diameter)
a)
Particle spectrum
Respirable fraction
b)
Aerosol output
Aerosol output rate
Aerosol data according to EN 13544-1; Alternative test method in accordance with Annex CC.
Nebulised medium: ARIKAYCE 8.4 ml.
a) Measurement with Next Generation Pharmaceutical Impactor (NGI) at 23 °C and 50% relative
humidity. Inspiratory flow: 15 l/min.
b) Measurement with breathing simulator at 23 °C and 50% relative humidity. 500 ml breathing volume,
breathing frequency 15 cycles/minute, sinusoidal breathing pattern, inhalation/exhalation ratio 1:1
(with adults, may vary in children).
Particle size distribution according to EN 13544-1
52
a)
< 5 μm
b)
Membrane nebuliser
approx. 55 g
5.0 cm x 6.7 cm x 14.5 cm
8.4 ml
4.4 μm
4.3 μm to 4.8 μm
55.7% +/- 3.6%
4.46 ml
0.35 ml/min
Lamira
®
- 2020-08