35
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The [KD-735] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [KD-735] should assure
that it is used in such an environment.
IMMUNITY test
Radiated RF
IEC 61000-4-3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the [KD-735] is used
exceeds the applicable RF compliance level above, the [KD-735] should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
the [KD-735].
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
IEC 60601
Compliance
test level
level
3 V/m
3 V/m
80 MHz
to 2.5 GHz
ENGLISH
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the [KD-735], including cables,
than the recommended separation
distance calculated from the equa-
tion applicable to the frequency of
the transmitter.
Recommended separation
distance:
d = 1.2 √ P
d = 1.2 √ P
80 MHz to 800 MHz
d = 2.3 √ P
800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
a
survey,
should be less than the
compliance level in each
b
frequency range.
Interference may occur
in the vicinity of equipment
marked with the following
symbol: