Foramen Needle Test Simulation
26 26
8
5
6
1
9
4
7
2
3
Grey (unassigned): Stimulation level is zero
•
Green ("good"): Stimulation level between 0 and 2 mA .
•
Placement is acceptable .
Yellow ("not ideal"): Stimulation level between 2 and 3 mA .
•
Axonics recommends adjusting the needle placement .
Red ("not recommended"): Stimulation level between 3
•
and 12 .5 mA . Axonics strongly recommends adjusting the
needle placement .
Press ( 8 ) in the upper right to move on to place the tined
•
lead .
For a PNE lead implant, this will say "PNE Lead Placement"
•
For a tined lead implant, this will say "Tined Lead
•
Placement"
To exit back to the Home screen, press "Exit" ( 9 )
•