Programming the Stimulator Stimulation Settings
Electrode Configuration
1
The CP provides the ability to use automatically generated
recommended electrode configurations or to manually set which
2
electrodes are active and inactive during stimulation .
Electrode Recommendations
The Programming screen displays up to four recommended
electrode configurations ( 2 ) . These recommendations are
intended to provide directional guidance regarding which
electrode configurations to test . For tined lead stimulation,
the recommendations are generated based on the stimulation
thresholds and motor and sensory responses recorded
during tined lead placement or subsequent redefining of the
thresholds . Ultimately, electrode configuration selection should
be based on patient comfort and symptom reduction . To set the
electrode configuration to a recommended configuration, press
the image of the recommendation .
Note: PNE lead electrode recommendations reflect all available electrode
configurations .
Note: If tined lead thresholds are not saved to the Stimulator, no electrode
recommendations are provided .
49