Setting Up a New Stimulator

1
2
4
5
3
There are several steps to set up a new Stimulator .
A prompt will appear when a new Stimulator is connected
if the trial type has not yet been assigned to the Stimulator .
Select the type of trial and the Trial Stimulator will be
configured appropriately .
Input the serial number of the patient's Remote Control in the
Remote Control ID field ( 1 ) . This pairs the Remote Control to
the Stimulator, allowing the Stimulator to receive commands
from the Remote Control .
Note: The Stimulator can only communicate with the Remote Control that is
entered in this field .
Enter the patient identifier associated with this device into the
Patient ID field ( 2 ) .
Enter the date of the Stimulator implant procedure in the
Device Implant Date field .
These fields, including the Patient ID and the Remote Control
serial number can be updated at any time as needed .