Error Messages - Hach SIP 10 Manual Básico Del Usuario

Ocultar thumbs Ver también para SIP 10:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Figure 2 Pump tubing change
Troubleshooting

Error Messages

Potential SIP 10 errors are displayed by the photometer.
Table 3 Error Messages
Error displayed
Cause
Sipper module not
Connection between
connected.
SIP 10 and photometer
disrupted.
Please check cable.
Please check sipper
Pump tubing not
and tube.
correctly inserted.
Resolution
Check the USB cable.
The cable length must not
exceed 1 m (3.3 ft) and the
cable must be directly
connected to the photometer.
Remove any other devices that
are connected where
necessary.
Loosen the cover and position
again. Where required, run the
pump briefly before the lever is
moved to make sure that the
tube is correctly positioned
around the rollers.
English 11

Publicidad

Tabla de contenido
loading

Tabla de contenido