6.1

Calibration

CAUTION!
The gradient must be calibrated
after every manipulation on the
sensor (e.g. electrolyte replace‐
ment, emery treatment, etc)!
The gradient has to be re-cali‐
brated at regular intervals to
ensure the perfect operation of
the sensor ! If there are no other
regulations then the sensor only
needs to be calibrated every 3-4
weeks for use in drinking water
and swimming pool water.
Avoid air bubbles in the sample
water! Air bubbles, which adhere
to the sensor can result in too low
a reading and thus lead to dan‐
gerous over-metering.
Please note the pertinent national
guidelines for calibration inter‐
vals!
Prerequisites
The sensor is ready (wait for run-
in time).
Constant flow rate at the flow
gauge
Constant temperature of the
sample water
Each time after either installing or
removing the sensor, wait until
the run-in time expires and a con‐
stant reading is achieved, but at
least 15 minutes to avoid drifts
caused by the temperature equal‐
isation.
No concentration fluctuations of
the feed chemical in the sample
water
Constant pH value within the per‐
missible range
The sample must be taken at the
place the sensor is installed