21
110kg).
The device may not work for all patients. If you are
unable to achieve stable readings, discontinue use.
The update period of data is less than 5 seconds,
which is changeable according to different individual
pulse rate.
If some abnormal conditions appear on the screen
during test process, pull out the finger and reinsert to
restore normal use.
The device has normal useful life for three years since
the first electrified use.
The hanging rope attached the product is made from
Non- allergy material, if particular group are sensitive
to the hanging rope, stop using it. In addition, pay
attention to the use of the hanging rope , do not wear
it around the neck avoiding cause harm to the patient.
The instrument dose not have low-voltage alarm
function, it only shows the low-voltage, please
change the battery when the battery energy is used
out.
When the parameter is particularly, the instrument
dose not have alarm function. Do not use the device
in situations where alarms are required.
Batteries must be removed if the device is going to be
stored for more than one month, or else batteries may
leak.
A flexible circuit connects the two parts of the
device. Do not twist or pull on the connection.
1.4 Indication for Use
The Fingertip Pulse Oximeter is a non-invasive device
intended for the spot-check of oxygen saturation of
arterial hemoglobin (SpO
and pediatric patients in home and hospital environ-
ments (including clinical use in internist/surgery, anes-
thesia, intensive care ect.).
This device is not intended for continuous monitoring.
ENGLISH
) and the pulse rate of adult
2