will be more sensitive to interference.
2. For those with a substantial amount of staining dilu-
tion drug (such as methylene blue, indigo green and
acid indigo blue), or carbon monoxide hemoglobin
(COHb), or methionine (Me+Hb) or thiosalicylic he-
moglobin, and some with icterus problem, the SpO
determination by this monitor may be inaccurate.
3. The drugs like dopamine, procaine, prilocaine, li-
docaine and butacaine may also be a major factor
blamed for serious error of SpO
4. As the SpO
judgement of anemic anoxia and toxic anoxia, some
patients with serious anemia may also report good
SpO
measurement.
2

4 TECHNICAL SPECIFICATIONS

1. Display Format: Digital tube Display
SpO
Measuring Range: 0% - 100%
2
Pulse Rate Measuring Range: 30 bpm - 250 bpm
Pulse Intensity Display: columniation display
2. Power Requirements: 2 ×1.5V AAA alkaline battery,
adaptable range: 2.6V~3.6V
3. Power Consumption: Smaller than 25 mA
4. Resolution: 1% for SpO
5. Measurement Accuracy: ±2% in stage of 70%-100%
SpO
, and meaningless when stage being smaller than
2
70%. ±2 bpm or ±2% (select larger) for Pulse Rate.
6. Measurement Performance in Weak Filling
Condition: SpO
rectly when pulse-filling ratio is 0.4%. SpO
±4%, pulse rate error is ±2 bpm or ±2% (select larger).
7. Resistance to surrounding light: The deviation be-
tween the value measured in the condition of man-
made light or indoor natural light and that of darkroom
is less than ±1%.
8. It is equipped with a function switch. The Oximeter can
be powered off in case no finger is in the device.
value serves as a reference value for
2
2
and pulse rate can be shown cor-
2
ENGLISH
measure.
2
and 1 bpm for Pulse Rate
2
error is
2