Message
Cannot
The unit cannot reach the target flow setting.
reach target
Check the heated breathing tube or patient interface for blockage.
Check whether the target flow setting is too high for the patient interface
flow
being used (refer to "Setting up myAIRVO 2" - "Select Patient Interface").
You will be prompted for acknowledgement.
WARNINGS
• The oxygen concentration delivered to the patient can be affected by
changes to the flow setting. Adjust the level of oxygen from the oxygen
source as necessary.
Check water
The chamber has run out of water.
If using the HC360 reusable chamber: Remove the chamber and refill.
If using the MR290 auto-fill chamber: When a chamber runs dry, the chamber
float may be damaged. Replace the chamber and water bag.
To ensure continual humidification, always ensure that the water chamber
and/or water bag are not allowed to run out of water.
Cannot
The unit cannot reach the target temperature setting.
reach target
You will be prompted for acknowledgement.
temperature
The most likely cause for this is that the unit is operating at a high flow rate in
low ambient conditions. Consider increasing ambient conditions to match the
recommended operating conditions or decreasing the target flow setting.
WARNINGS
• The oxygen concentration delivered to the patient can be affected by
changes to the flow setting. Adjust the level of oxygen from the oxygen
source as necessary.
[Power out]
The unit has been disconnected from the mains/utility power socket.
No visual alarm. The auditory alarm will sound for 120 seconds. If power is
reconnected in this time, the unit will automatically restart.-*
WARNINGS
• Appropriate patient monitoring must be used at all times. Loss of therapy
will occur if power is lost.
ALARM LIMITS
Most alarm limits are pre-programmed. The exceptions are listed below. These alarm limits may be
changed to other values by authorized personnel. Changes will be preserved during or after any power
loss.
Alarm
Factory-set
condition
alarm limit
O
too low
2
O
too high
2
WARNINGS
• A hazard can exist if different alarm presets are used on different units within any single area, eg. long term care facility
• Alarm limits set to extreme values can render the alarm system useless.
CHECKING ALARM SYSTEM FUNCTIONALITY
The functionality of the alarm system can be checked at any time when the unit is turned on.
Remove the heated breathing tube. You should see the "Check tube" visual alarm signal and hear the
auditory alarm signal. If either alarm signal is absent, do not use the unit. Contact your Fisher & Paykel
Healthcare representative.
AUDITORY INFORMATION SIGNALS
In addition to auditory alarm signals, auditory information signals are provided. These are described below.
Melody
Ascending sequence of 5 tones
Ascending sequence of 3 tones
Decending scale of 3 tones (within 2 seconds)
Single tone every 5 seconds
Single tone every 30 seconds
Meaning
Possible preset
values
21% O
21 or 25% O
2
30 – 95% O
90% O
in 5% increments
2
2
2
The "Ready for use" symbol has appeared
Activation/deactivation of Junior Mode
Drying Mode has been activated
Measured oxygen level ≥ 33% at turn-off
Measured oxygen level > 95%
A – 12
Affects
delivery
of:
Oxygen
Humidity
Humidity
Oxygen,
humidity.
Meaning
Delays
< 120
seconds
< 30
minutes
30 +/- 3
minutes
< 5 seconds