2. SETTING UP myAIRVO 2
H
O
2
1. BEFORE YOU BEGIN
Place the unit on a low shelf or near the floor beside your bed. It must
be placed below head height and flat. Position the device so the power
cord connection to the power supply is easily accessible and able to be
disconnected.
2. INSTALL WATER CHAMBER
IF USING A HC360 REUSABLE WATER CHAMBER:
With the aid of the supplied funnel, fill the chamber with enough distilled
water for the period of use, but never above the 560 mL maximum water
level line.
(HC360 reusable water chamber, at 37 °C target temperature)
L/min
2
5
10
hrs
106
42
21
WARNINGS
To avoid burns:
• Do not fill the water chamber with hot water.
To avoid electric shock:
• Always remove the water chamber to fill it and always fill with enough water to
prevent it running out.
Fit the water chamber to the unit by pressing down the finger guard and
sliding the chamber on, carefully aligning with the blue chamber port ends.
Push the chamber on firmly until the finger guard clicks into place.
Go to Step 3, "Install Heated Breathing Tube", below.
CAUTION
Adding substances other than water can adversely affect the humidifier and delivered
therapy.
IF USING THE MR290 AUTO-FILL WATER CHAMBER:
Remove the blue port caps from the chamber by pulling the tear tab
upwards then remove the bracket holding the water supply tube. Fit the
supplied adapter over the two vertical ports on the chamber and push on
fully then clip the water supply tube into position.
Fit the MR290 chamber as described above for the HC360 chamber.
Attach the water bag to the hanging bracket at least 20 cm (8") above
the unit, and push the bag spike into the fitting at the bottom of the bag.
Open the vent cap on the side of the bag spike. The chamber will now
automatically fill to the required level and maintain that level until the
water bag is empty. Use only distilled water and always ensure sufficient
water is in the water bag to prevent it from running out.
A – 4
Flow setting vs usage time
15
20
25
30
35
14
10
8
7
6
40
45
50
55
60
5
4
4
3
3