Power-save operation
The time after which the QuikRead go instrument
closes the lid – without entering sleep mode – can
be adjusted from Power save > Close lid only; adjust
the delay value with the arrow buttons. The instru-
ment will close the lid if it hasn't been used for the
amount of time specified, without entering sleep
mode (activating the hibernation function). This
mode does not disrupt any LIS/HIS connection.
The time after which the QuikRead go instrument
enters sleep mode can be adjusted from Power save
> Full standby; adjust the delay value with the arrow
buttons. The instrument goes from idle mode into
sleep mode if it hasn't been used for the amount
of time indicated. Entering sleep mode breaks any
active LIS/HIS connection.
Accept the settings with OK, or reject with Cancel.
Saving changes to personal settings
After all desired adjustments to personal settings
have been made, choose Save.
Saving setting to profiles for further use
From the main menu screen, choose Profiles.
Choose Save as profile, and select either use of a
new profile or a profile you'd like to copy and then
modify. Supply a name for the profile, and accept
with OK.
20
PRINCIPLES OF OPERATION

Measurement flow

With the measurement flow settings, the operator
can adjust laboratory- or work-routine-oriented set-
tings such as operator and patient IDs, printing, LIS
transfer, and some test-specific parameters.
These settings can be selected for temporary use
(use until power-down) with Save after you make
the desired changes.
For further use, the settings need to be saved to a
profile. Settings for continuous use should be con-
figured with the set-up wizard.
Operator ID
... is an identifier of the user.
Operator ID OFF: The instrument does not re-
quest an Operator ID.
Operator ID ON: An Operator ID has to be given
before every sample measurement. That ID is
paired with the test result.
Operator ID ON + Propose previous: The instru-
ment suggests a previous ID for use, but a dif-
ferent ID can be supplied.
Patient ID
... provides identification of the patient sample.
Patient ID OFF: The instrument does not request
a Patient ID.
Patient ID ON: A patient ID has to be supplied
before every measurement, and that ID is pre-
sented with the test result.
Test parameters
The unit of results can be chosen here. The change
requires the administrator's password, which is QR-
...
GOSET. Choose
and the options will be shown.
To make settings permanent, they should be config-
ured with the set-up wizard.
Printing
Printing OFF: The instrument does not prompt
for printing. It is nevertheless possible to print
the result if one chooses Print from the Measure
/ Result screen.
Printing ON: After lifting the cuvette, the instru-
ment asks "Print current result?". Accept print-
ing by choosing Yes. Cancel printing by choos-
ing No.
Printing ON + Automatic: The instrument prints
every measurement result automatically.
LIS transfer
LIS transfer OFF: The instrument does not send
the results to the Laboratory Information System.
LIS transfer ON: After printing, if activated, the
instrument asks: "Send result to LIS?" Accept
sending by choosing Accept. Reject by choosing
Reject.
LIS transfer ON + Automatic: The instrument
sends the measurement result automatically to
the LIS.