Lifting/transporting the instrument
When lifting or transporting the QuikRead go
instrument, always handle it carefully. At the back of
the instrument is a handle, enabling lifting with one
hand (Image 4).
A recess is added to the instrument side surfaces to
help the grasp (Image 5).
Image 4
Lifting the instrument with one hand.
8
INSTALLATION PROCEDURES AND REQUIREMENTS
Location and environment
During use
The instrument should be placed on a flat, clean,
horizontal surface, and the following restrictions
should be observed:
Image 5
Lifting the instrument with two hands.
●
Altitude up to 6,500 feet (2,000 m)
●
Ambient temperature between 59 °F (15 °C)
and 95 °F (35 °C)
●
Maximum relative humidity 80% for tempera-
tures up to 87 °F (31 °C), decreasing linearly to
67% at 95 °F (35 °C)
●
Power-supply voltage fluctuations of no more
than ±10% of the nominal voltage
●
Installation category II (2,500 V transient)
●
Do not expose the instrument to direct sunlight
●
Do not use this device in close proximity to
sources of strong electromagnetic radiation (e.g.
unshielded intentional RF sources), as these may
interfere with the proper operation
●
Do not move the instrument while it is perform-
ing measurements
●
Noise level generated is <85 dB (A)
●
Environmental limit of Pollution Degree 2
During transport and storage
●
The ambient temperature must be between 36
°F (2 °C) and 95 F (35 °C)
●
Protect from rain and humidity
●
Handle the instrument with care