Lifting/transporting the instrument

When lifting or transporting the QuikRead go
instrument, always handle it carefully. At the back of
the instrument is a handle, enabling lifting with one
hand (Image 4).
A recess is added to the instrument side surfaces to
help the grasp (Image 5).
Image 4
Lifting the instrument with one hand.
8
INSTALLATION PROCEDURES AND REQUIREMENTS

Location and environment

During use

The instrument should be placed on a flat, clean,
horizontal surface, and the following restrictions
should be observed:
Image 5
Lifting the instrument with two hands.
Altitude up to 6,500 feet (2,000 m)
Ambient temperature between 59 °F (15 °C)
and 95 °F (35 °C)
Maximum relative humidity 80% for tempera-
tures up to 87 °F (31 °C), decreasing linearly to
67% at 95 °F (35 °C)
Power-supply voltage fluctuations of no more
than ±10% of the nominal voltage
Installation category II (2,500 V transient)
Do not expose the instrument to direct sunlight
Do not use this device in close proximity to
sources of strong electromagnetic radiation (e.g.
unshielded intentional RF sources), as these may
interfere with the proper operation
Do not move the instrument while it is perform-
ing measurements
Noise level generated is <85 dB (A)
Environmental limit of Pollution Degree 2

During transport and storage

The ambient temperature must be between 36
°F (2 °C) and 95 F (35 °C)
Protect from rain and humidity
Handle the instrument with care