Quality Control - Orion QuikRead go Instrucciones De Uso

Ocultar thumbs Ver también para QuikRead go:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

2.
Drop a cuvette into the measurement well in
the right position. The barcode of the cuvette
must be facing towards you (see Image 19).
Note: do not insert your finger or anything else
into the measurement well.
3.
The lid closes, and the instrument starts the
measurement.
4.
After the test is completed, the result is shown
on the display and the cuvette is raised for re-
moval.
5.
Remove the cuvette. The result disappears from
the display. It can be displayed again via See
previous result.
6.
If you wish to make another measurement,
insert a new cuvette into the measurement well.
Choosing Cancel takes you to the main menu.

Quality control

QuikRead go has a separate result history file for
quality-control samples. Quality-control samples are
measured in the same way as normal samples, but
the results are stored in a separate result file. To start
a quality control measurement, choose Quality con-
trol in the main menu and follow the instructions
on the screen.
26
OPERATING INSTRUCTIONS
Image 19
Drop a cuvette into the measurement well
with the barcode facing towards you.