7-2 Operation
7-2-1 Alternate operation
) When the ON/OFF Key is pressed briefly, the
Motor Handpiece starts rotation.
2) When the ON/OFF Key is pressed again, the
rotation stops.
7-2-2 Intermittent operation
) Press and hold down the ON/OFF Key for one
second or longer.
2) While the Key is held down, the Motor Handpiece
rotates.
3) When the ON/OFF key is released, the rotation stops.
For fine-tuning the rotation speed and the torque upper limit, press the SPEED Key and the TORQUE Key to
adjust the values.
When the treatment is over, press and hold down the POWER Key for one second or longer to turn off power.
8
Various Functions
8-1 Memory Function
The values (speed, torque, gear ratio, auto reverse mode) set as required can be saved under a chosen
program.
) Press the PROGRAM Key to select a Program No. for saving the setting into memory.
2) Set the speed, torque, gear ratio, auto reverse mode as required.
3) Press and hold down the MEMORY Key for one second or longer. When the alarm sounds, the setting has
been saved in the selected program.
NOTICE
•
A program cannot be saved while the Motor Handpiece is rotating.
•
The Rotation Modes cannot be saved in a program. Also, when the rotation mode is set at
"R (reverse)", the setting cannot be saved.
•
When the PROGRAM Key is pressed and held down, Program No. changes continuously.
•
The MEMORY Key needs to be pressed and held down for two seconds or longer to save
the program. If the MEMORY Key is not pressed long enough, the setting is not saved in the
memory. If the Program No. is changed using the PROGRAM Key before the MEMORY Key is
pressed, the setting remains the original saved values (cancel function).
8-2 Restoring the Default Setting of Programs
The factory setting can be restored as follows.
) Turn OFF the power.
2) While pressing the AUTO REVERSE Key and the MEMORY Key together, press and hold down the POWER
key for one second or longer.
3) When the alarm sounds and "
is displayed, press the MEMORY Key.
4) When "
" appears on the LCD Panel, the default setting has been restored.
" has appeared on the LCD Panel, release both keys. Then while "
8
ON/OFF Key
Fig. 4
"