Storage; Operating Environment; Travel - Orthofix SPINAL STIM 5212CE Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Storage

When moving the SpinalStim device from very cold or very hot storage areas
(like your car), wait at least an hour to use or charge the device. The device
requires time to return to operating temperature.
Unpacked Storage:
Temperature Range:
• -25°C to 5°C
• 5°C to 35°C at up to 90% relative humidity, non-condensing
• 35°C to 60°C at a water vapor pressure up to 50 hPa
Packed Storage, Shipping and Transport:
Temperature Range: within -40°C to 60°C
• Between 10-100% relative humidity
• Including condensation at pressures between 500 hPa and 1060 hPa

Operating Environment:

Temperature Range: within 5°C to 40°C
• 15-90% relative humidity, non-condensing but not requiring a water
vapor pressure greater than 50 hPa
• 700-1060 hPa
The SpinalStim device is designed for a storage life of twelve months plus one
year of usage.

Travel

When traveling by air, it is recommended to pack the SpinalStim device with
checked luggage. If taken onboard the airplane, it should be turned off
when passing through security screening equipment, as the device could be
damaged. The SpinalStim device instruction manual should be taken with you
to quickly and easily identify the device for security personnel. Do not wear or
operate the SpinalStim device while onboard the airplane.
12

Publicidad

Tabla de contenido
loading

Tabla de contenido