S Series Ultrasound System
User Guide Errata P12397-04
Attaching and removing USB devices
If you attach a USB device to the system, wait at least 30 seconds before removing it or attaching another USB
device. Similarly, if you remove a USB device from the system, wait at least 30 seconds before reattaching it or
attaching another USB device.
Attaching or removing USB devices (for example, a USB printer or a USB storage device) too quickly may cause the
system to stop responding and to display a blue screen with the Maintenance
and Attention
icons.
Patient report
If you transfer or export an exam from the S Series ultrasound system, the following measurements
do not appear in the exported patient report. If you need to record them, save the report as an image
(press Save) before transferring or exporting, or write them manually into the patient record.
• GYN exam: Follicle
• Cardiac exam: Automatically calculated CO values (CO, SV, CI, SI, VTI, Vmax, Vmean, PGmax, PGmean, LVOT D,
LVOT area)
Language selection
If you need to change the Language selection on the Presets setup page, make sure to first archive, transfer, or
export all saved patient information you want to keep. The system may not correctly maintain the saved patient
information after the Language selection is changed.
Video Mode selection
If you change the Video Mode selection (NTSC or PAL) on the Connectivity setup page, restart the system.
Changing the Video Mode selection can disable the USB ports until the system is restarted.
Acoustic output tables
The following tables provide updated acoustic output information.