5.1 Energy-saver mode (Auto power off)

If work is interrupted for more than three minutes (factory setting), the instrument automatically
switches into standby mode. The display value is saved and reappears in the display after switching
back on manually. The user can adjust the amount of idle time before automatically switching off (see
Auto power off (APO mode), p. 73).

5.2 Pause function

If the instrument has not been completely primed, air bubbles appear in the titrating tube during titra-
tion. To prime the instrument during titration, liquid can be dispensed into a different receptacle us-
ing the Pause function without the display value changing.
Pause
1 x
9974 69
a.
Press the Pause button one time.
ð
The Pause signal blinks.
b.
Prime the instrument, dispense liquid, etc. See Prim-
ing, p. 58.
c.
To end the Pause function, briefly press the Pause but-
ton again.
ð
The Pause signal disappears.
d.
Continue titration
5 Titrating
61
Operating manual