jbc advanced AM 6500 Serie Manual De Instrucciones página 10

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Exchanging the resistor from the heater.
This step has to be performed when the heater
This step has to be performed when the heater
This step has to be performed when the heater
This step has to be performed when the heater
This step has to be performed when the heater
is cold
is cold. . . . .
is cold
is cold
is cold
1. Untighten the screws.
2. Separate the resistor from the heater's cable.
3. Connect the new resistor, pushing it's extreme.
4. Fix the screws tightly to avoid air-loss which
could reduce the resistor's lifetime.
OPERATING INCIDENTS
The suction cup does not adhere to the
component.
Deficient aspiration, Vacuum.
1 Verify if the suction cup is well placed and in
perfect condition.
2 Check the incoming air filter in the interior of
the station and replace it if dirty or obstructed.
Filter
Ref. 0861800
Possible errors
In the case of malfunction, the control unit
interrupts the connection.
Following a list of the most common
malfunctions:
- - - - - Power failure.
Check for blown fuses.
- The temperature will not rise.
Possible causes:
· Heating element open.
· In case of a long low-power period.
- No reading from the thermocouple.
Possible cause: open thermocouple.
- Insufficient air flow which causes an excesive
rise of the heating temperature.
Before recuperating this type of error you must
wait until the temperature goes down.
Possible causes: leaking or blocked air
conducts or faulty self-contained air pump.
- Faulty reading of the rotationsmeter of the self-
contained air pump.
Possible causes: air pump damaged or faulty
function of the optical sensor circuit.
To recuperate any of these errors actuate the
general switch at the back of the station, the
pedal should not be pushed at this moment.
9
ENGLISH

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Advanced serieAdvanced ar 5800

Tabla de contenido