MANUAL DE INSTRUCCIONES CODIGO 80196 REV B
Use the key ESC
to return to pre-
vious step.
Press ENTER to entry into menu
OPERATION
OPERATION.
PROGRAM
01
XXXXXXXX
PROGRAM
02
XXXXXXXX
Use key ENTER
to run the extraction
program.
J
.P. SELECTA s.a.u.
Octubre 2019
To run a program:
•
Open the water tap for coolant water.
•
Select the menu «OPERATION» to run an stored program. See 5.2.
- Bear in mind that program number "0" is only for checking purposes.
- After «OPERATION» is selected, the first program shown on the display is the
last program used or edited.
- To start a program, press key ENTER/START and could be stopped at any time
by pressing key ESC/STOP.
•
The program initiates the «BOILING» step, so place the thimbles on the low posi-
tion (See 5.5). Once the «BOILING» step finished, the unit makes some BEEP and
the display flashes.
Do not let the unit completely unattended.
Pay attention to the end of each step to change, at time,
•
The unit jumps directly to «RINSING» step. Move the thimble to the up position.
Once the «RINSING» time finishes, the unit makes some BEEP and the display flashes.
•
The unit jumps directly to «RECOVERY» step. Close the PTFE manual valves for
solvent recovery. Once the «RECOVERY» time finishes, the unit makes some BEEP
and the display flashes.
SOLVENT
DIETHILIC ETHER
Use keys (Up/Down) to select the program to run.
SOLVENT
ACETONE
Program step
BOILING
19:40
Remaining
time mm:ss
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso)
the extraction thimbles position.
Bo Ri Re TºC
10 10 10 80
Bo Ri Re TºC
10 10 10 140
Display when a program is running
Set temperature
Set time
20m t85
57,4
Actual tempe-
Automatically apears the parameters
for the last program used (edited or
run).
The display flashes between them.
Heating status:
1
0: Stopped
1: Active
rature
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 30