External Load Cell Settings _ Cal 2; Zero Point Calibration - Baxtran IKP Serie Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

IKP model
NPluS
Add
11. MR setting
DISPLAY
Nr rEC
LEAuE
Nr rEC
CIEAr

EXTERNAL LOAD CELL SETTINGS _ CAL 2

Press
to browse the menu options. Use the numeric keys to input data and press
settings.
1. Setup B scale capacity
DISPLAY
b.LoAd
XXXXXXX
GrAN
2. Setup B scale interval
DISPLAY
b.d =
XXXXX
GrAN
3. Setup B scale stable control level
DISPLAY
b.StAbL
X
1-10
Minimal level is 1, allowing the fastest stabilization time but with slowest noise filter.
Maximum level is 9, allowing best noise filter but with slowest stabilization time.

4. Zero point calibration

DISPLAY
XXXXXX
0
b.Pnt.0
Make sure the weighing pan is empty. When the AD value becomes stable, press
Press
to add the weighing result only.
DESCRIPTION
Press
to show the total weighing result.
Pressing the key again will not clear the total in memory.
Press
to show the total weighing result and press the key again to clear the memory.
Note: When the LP-50 printer is connected to the scale, pressing
time will clear the memory and print out the total counting result. The serial
connection in CAL 3 should be set to LP-50.
DESCRIPTION
Setup the capacity of B scale by gram.
If input 0 here, B scale will be disabled.
DESCRIPTION
Setup the scale interval (d) of B scale by gram.
Min. d is 1 gram.
Max. d is 65535 gram.
DESCRIPTION
Setup the stable control level.
X: stable control level 1-10.
The default stable level is set to 3.
DESCRIPTION
The first row shows AD Value, the second row shows "0", and the third row shows
"b.Pnt.0"
User's Guide