User's Guide
When the AD value becomes stable, press
LINEARITY CALIBRATION PROCEDURE _ CAL 5
Note: This section can only be operated by engineers.
Please remove the jumper JP3 switch OFF before you start the calibration process.
In Calibration menu, press
1. SEL maximum capacity
DISPLAY
Kg
LoAd
03
LoAd
06
LoAd
15
LoAd
30
LoAd
50
Note: the capacity will displayed in kg or lb according to the preset weighing unit in CAL 1.
2. Zero point calibration
DISPLAY
XXXXXX
Makes sure the weighing pan is empty. When the AD value becomes stable, press
3. Calibration by 1/3 of the maximum capacity
DISPLAY
g
lb/1000
XXXXX
XXXXX
2000
Pnt 1
Put on poise of the 1/3 weight of the full load. When the AD value becomes stable, press
4. Calibration by 2/3 of the maximum capacity
DISPLAY
g
lb/1000
XXXXX
XXXXX
4000
Pnt 2
Put on poise of the 2/3 weight of the full load. When the AD value becomes stable, press
42
to toggle among the menu options, and
DESCRIPTIONS
Lb
Max. capacity displayed varies with preset wt unit
LoAd
Set maximum capacity as 3Kg/6lb
06
LoAd
Set maximum capacity as 6Kg/12lb
12
LoAd
Set maximum capacity as 15Kg/30lb
30
LoAd
Set maximum capacity as 30Kg/60lb
60
LoAd
Set maximum capacity as 50Kg/100lb
100
DESCRIPTION
0
The first row shows AD value, the second row shows "0" and the third row shows "Pnt.0".
Pnt.0
DESCRIPTION
Calibration unit displayed varies with preset weight unit
The first row shows AD value, the second row shows the weight of 1/3 of the maximum
1000
(Unit: gram or lb/1000) and the third row shows "Pnt 1".
Pnt1
DESCRIPTION
Calibration unit displayed varies with preset weight unit
The first row shows AD value, the second row shows the weight of 2/3 of the maximum
2000
(Unit: gram or lb/1000) and the third row shows "Pnt 2".
Pnt2
and the calibration is done.
IKP model
to confirm.
.
.
.