Area C is the area of connection mode selection and connection status indication.
When the software is opened, the default is WIFI connection. This icon indicates that the current connection is WIFI connection.
You can switch to serial connection mode by clicking the icon button.
Serial Port Connection Instructions, you can switch to WIFI Connection Mode by clicking on this icon.
Text is used to display the current connection mode. The serial number of the current connected
torque tester is displayed at the combination of numbers and letters.
The alternating flickering of green and grey dots indicates that the current connection has been
established, and the constant brightness of grey dots indicates that the current connection has not been established. Text display shows
the data interaction between the torque tester and the software. There are two kinds of situations, normal and abnormal.
Area D is the real-time and peak torque feedback from the torque tester.
Real-time Torque is Tracking Torque.
The peak torque is the first peak torque of the wrench, and if the peak torque is qualified (whether the difference between the
set torque and the set torque is within the percentage deviation of the set tolerance), the peak torque is shown in green and the
unqualified is shown in red (see the example below).
Area E is the display and preservation area of multiple test results.
The evaluation results are included in the status bar. If all the test data are qualified, it will be OK. If one set of values is not qualified, it
will be NOK.When OK, the status bar is green and the evaluation result is OK. When NOK, the status bar is red and the evaluation result is
NOK.
The maximum, minimum and average values are the maximum, minimum and average results of multi- group data statistics in the
general mode (displayed as 0 in the verification mode), and the residual effective values are the number of remaining preloads in the
verification mode (set in the system management, set as 0 in the general mode).
Each row below has five historical peaks, up to 25.