TC-MIDI
6 Maintenance
Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator's manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.
7 Service and Repair
Please refer to the original operator's manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.
7.1 Error messages on the display
If an error occurs, the display will read "Err". Press the "
button to show the error number(s).
Error messages will appear until the unit has been restarted
or the error is cleared using the "Func" button. It can only be
cleared if the cause for the error has been corrected.
Causes / Action: The following is a list of the most common
causes and actions for the respective error. If the actions listed
do not resolve the problem, please contact Service.
Problem / Mal-
Possible cause
function
No dis-
• No voltage
play
• Loose connect-
ing cable
• Display defect-
ive
D1.02
(The software ver-
sion for the display
(perman-
will appear).
ent)
• Not communic-
ating with the
controller
Error
• An error has oc-
curred
Error 01
• Controller mal-
function
Error 03
• µC fault / MCP2 • Contact service
Error 04
• EEPROM error
Error 22
• Moisture de-
tector 1 cable
break
Error 32
• Moisture de-
tector 2 cable
break
BX440016 ◦ 12/2016
"
Action
• Check the supply
cable
• Check fuse
• Check connec-
tions
• Check connec-
tions
• Read the error
number as de-
scribed above
• Clear error (tem-
porary fault)
• Disconnect from
power for approx.
5 s
• Contact service
• Contact service
• Check moisture
detector line
• Check moisture
detector
• Check moisture
detector line
• Check moisture
detector
Error 40
• General error
temperature
sensor 1
Error 41
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 1
Error 42
• Excess temper-
ature / short-
circuit temper-
ature sensor 1
Error 43
• Measurement
fluctuation
temperature
sensor 1
• µC fault
Error 50
• General error
temperature
sensor 2
Error 51
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 2
Error 52
• Excess temper-
ature / short-
circuit temper-
ature sensor 2
Error 53
• Measurement
fluctuation
temperature
sensor 2
• µC fault
Bühler Technologies GmbH
• Sensor possibly
defective
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Sensor possibly
defective
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
15