Error Messages On The Display - Bühler technologies TC-Double Guía Rápida

Ocultar thumbs Ver también para TC-Double:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 26
TC-Double (+)
6 Maintenance
Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator's manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.
7 Service and Repair
Please refer to the original operator's manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.

7.1 Error messages on the display

If an error occurs, the display will read "Err". Press the "
button to show the error number(s).
Error messages will appear until the unit has been restarted
or the error is cleared using the "Func" button. It can only be
cleared if the cause for the error has been corrected.
Causes / Action: The following is a list of the most common
causes and actions for the respective error. If the actions listed
do not resolve the problem, please contact Service.
Problem / Malfunc-
Possible cause
tion
No display
• No voltage
• Loose connect-
ing cable
• Display defect-
ive
D1.02
(The software ver-
sion for the display
(perman-
will appear).
ent)
• Not communic-
ating with the
controller
Error
• An error has oc-
curred
Error 01
• Controller mal-
function
• Malfunction
controller 2
Error 03
• Microcontroller
Fault / MCP2
Error 04
• EEPROM error
16
Bühler Technologies GmbH
"
Action
• Check the sup-
ply cable
• Check fuse
• Check connec-
tions
• Check connec-
tions
• Read the error
number as de-
scribed above
• Clear error
(temporary
fault)
• Disconnect
from power for
approx. 5 s
• Contact service
• Clear error
(temporary
fault)
• Disconnect
from power for
approx. 5 s
• Contact service
• Contact service
• Contact service
• Fan malfunc-
tion / fan out of
order
Error 22
• Moisture de-
tector 1 cable
break
Error 32
• Moisture de-
tector 2 cable
break
Error 40
• General error
temperature
sensor 1
Error 41
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 1
Error 42
• Excess temper-
ature / short-
circuit temper-
ature sensor 1
Error 43
• Measurement
fluctuation
temperature
sensor 1
Error 50
• General error
temperature
sensor 2
Error 51
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 2
Error 52
• Excess temper-
ature / short-
circuit temper-
ature sensor 2
Error 53
• Measurement
fluctuation
temperature
sensor 2
• Check fan
(blocked?)
• Check connec-
tion
• Check moisture
detector line
• Check moisture
detector
• Check moisture
detector line
• Check moisture
detector
• Sensor possibly
defective
• Check temper-
ature sensor
connection
• Check temper-
ature sensor
connection
• Check temper-
ature sensor
connection
• Sensor possibly
defective
• Check temper-
ature sensor
connection
• Check temper-
ature sensor
connection
• Check temper-
ature sensor
connection
BX440021 ◦ 06/2017

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Tc-double+

Tabla de contenido