FujiFilm VISUALSONICS Vevo MD Manual Del Usuario página 9

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Did any part of the system or transducer come in contact with
broken skin, blood, or bodily fluids?
(semi-critical)
(non-critical)
Before getting started
Inspect the system and transducer to determine that it is free of any
l
unacceptable deterioration, such as corrosion, discoloration, pitting, or cracked
seals. If damage is evident, discontinue use, and contact FUJIFILM SonoSite
or your local representative.
Confirm that cleaning and disinfecting materials are appropriate for your
l
facility's use. FUJIFILM SonoSite tests disinfectants and disinfectant devices
for use with the system and transducer.
Verify expiration dates, concentration, and efficacy of chemicals (for example,
l
a chemical strip test). Disinfectants and cleaning methods listed are
recommended by FUJIFILM SonoSite for efficacy and material compatibility
with the products.
Ensure that the disinfectant type and the solution strength and duration are
l
appropriate for the equipment.
Cleaning and disinfecting guide
YES
OR
NO
Step A: Semi-critical contact
cleaning and disinfecting of
system and transducer
page 11
Step B: Non-critical contact
cleaning and disinfecting of
the system and transducer
page 18
Cleaning and Disinfecting - English
on
on
9

Publicidad

Tabla de contenido
loading

Tabla de contenido