Before Getting Started - FujiFilm VISUALSONICS Vevo MD Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Before getting started

Before cleaning and disinfecting:
Inspect the foot switch to determine that it is free of any unacceptable
l
deterioration, such as corrosion, discoloration, pitting, or cracked seals. If
damage is evident, discontinue use, and contact FUJIFILM SonoSite or your
local representative.
Confirm that cleaning and disinfecting materials are appropriate for your
l
facility's use. FUJIFILM SonoSite tests disinfectants and disinfectant devices
for use with the foot switch.
Verify expiration dates, concentration, and efficacy of chemicals (for example,
l
a chemical strip test). Disinfectants and cleaning methods listed are
recommended by FUJIFILM SonoSite for efficacy and material compatibility
with the products.
Ensure that the disinfectant type and the solution strength and duration are
l
appropriate for the equipment.
Follow manufacturer recommendations and local regulations, when preparing,
l
using, and disposing of chemicals.
Follow the disinfectant manufacturer's recommendations regarding appropriate
l
personal protective equipment (PPE), such as protective eye wear and gloves.
24
Cleaning and Disinfecting - English

Publicidad

Tabla de contenido
loading

Tabla de contenido