6.4.5. File Library
The device contains a file library with the following preset NiTi systems:
A. Reciprocating systems
! WaveOne
! RECIPROC
B. Continuous rotary systems
! Gates
! PathFile
! Protaper
! ProFile
! GT
Series X
®
! My Program (individual programs)
The manufacturer reserves the right to update the file library and the systems contained in it.
WARNINGS
! Follow the file manufacturer's instructions for use of endodontic files.
! The file system shown on the display must always match the file in use. This is of the
utmost importance in order to avoid misusing reciprocating files and continuous rotary files.
! Torque and speed values are subject to change by the file manufacturers without notice.
Therefore, the preset values in the library must be checked prior to use. Torque values
shown on the display are accurate and reliable only with WaveOne
contraangles properly maintained and lubricated.
6.5.
Operation
6.5.1. Switchon, Standby Mode and Switchoff
Switchon
Press the POWER key. All LEDs will light up. A welcome screen displays the current software
version.