Precautions - Dentsply Maillefer wave-one Manual Del Usuario

Ocultar thumbs Ver también para wave-one:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

! The battery charger must be supplied at a voltage ranging between: 100 – 240 V (+/­ 10%),
47­63 Hz. Use only original parts.
! If the BATTERY LED starts flashing red during use, connect the device immediately to the
battery charger, as the device may switch off.
! In order to ensure good battery life, it is recommended to always work on battery power
and recharging the battery only when it is fully discharged.
! Should any anomalies arise during operation, suspend work and contact your distributor.
! Do not connect an external PC storage unit (hard disk) to the WaveOne
socket. Never use a commercial male­male USB cable to connect the WaveOne
motor to a PC.
4.

Precautions

Read these safety precautions thoroughly prior to use. These precautions allow you to use the
product safely, preventing harm to you and others.
It is of the utmost importance that this manual is preserved for future consultation. The manual
must accompany the system in all cases of sale or other transfer in order that the new owner
may reference the precautions and warnings.
Gloves and a rubber dam are mandatory during the use of WaveOne
Refer to the WARNINGS chapter (see chapter 3) to verify any special care to exercise before
starting to use the complete device.
The manufacturer declines any responsibility in the case of:
! Use of the device for applications other than those specified in the instructions for use and
maintenance.
! Modifications or repairs performed by persons not authorized by the manufacturer.
! Connection of the motor to an electrical supply not in conformity with the provisions of the
IEC 364 standard.
! Use of non­original components or components other than those specified in the
STANDARD COMPONENTS chapter (see chapter 6.1).
! File breakage due to misuse.
! Accessories or device breakages due to the sterilization: none of the WaveOne
motor components are sterilizable (except for the contra­angle, see separate contra­angle
operating instructions).
endo motor USB
!
endo
!
endo motor.
!
endo
!
- 6 /141 -

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido