NOTE
Be aware that all individual settings, including the settings in My Program, will be deleted when
"reset to default parameters" has been carried out.
The acoustic signals settings are not affected: they remain with the last setting used before the
device was switched off.
ASR function can be modified for each file, but the setting will return to the default settings as
soon as a different file is selected.
6.8.
Maintenance
Routine Maintenance
! Inspect the micromotor cable at least once every six months. If any deterioration of the
sheathing is noticed, have the cable replaced by an authorized service center.
! Check for liquids or smoke exiting the device. In this case, remove the power supply from
the device immediately and contact an authorized service center.
! For optimal battery performance, replace the battery every 2 years.
! For the maintenance, e.g. lubrication of the WaveOne
refer to the separate contraangle operating instructions and check the additional warnings
below.
WARNINGS
! The contraangle should be lubricated after cleaning and disinfecting, but prior to
sterilization.
! When lubricating the contraangle, check that no lubricant penetrates the micromotor.
! If the contraangle is lubricated manually, please ensure excess oil is removed by
compressed air (blow for approx. 5 sec.) before the contraangle is being put back on the
micromotor.
! If the contraangle is lubricated automatically in a maintenance or lubrication device, please
follow the instructions for use of the device manufacturer carefully and ensure no excess oil
is left in the processed contraangle.
! Do not lubricate the micromotor for any reason. Lubricant may contaminate the micromotor
causing damage and result in unsafe operation. This will invalidate the warranty.
! Do not introduce any foreign objects into the micromotor shaft.
Special Maintenance
Contact your distributor for any special maintenance that may be required.
NOTE
Some operations carried out during repair could lead to deletion of personal settings such as
changed torque, speed or My Program settings.
(MAILLEFER 6:1) contraangle,
!
- 23 /141 -