The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
Test level
Immunity test
EN 606011 2
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
EN 6100042
Burst/Fast
±2 kV power
Transient
supply lines
EN 6100044
±1 kV differential
Surge
mode
EN 6100045
±2 kV common
mode
Voltage dips,
< 5 % UT
short interruptions
(> 95 % dip in UT)
and voltage
for 0,5 cycles
variations on
power supply
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
input lines
(30 % dip in UT)
EN 61000411
for 25 cycles
< 5 % UT
(> 95 % dip in UT)
for 5 seconds
Power frequency
magnetic field
3 A/m
EN 6100048
Immunity aspects
Compliance Level
Floors should be wood, concrete or ceramic tile.
± 6 kV contact
If floors are covered with synthetic material, the relative
± 8 kV air
humidity should be at least 30 %.
±2 kV power
Mains power quality should be that of a typical
supply lines
commercial or hospital environment.
±1 kV differential
mode
Mains power quality should be that of a typical
±2 kV common
commercial or hospital environment.
mode
< 5 % UT
(> 95 % dip in UT)
Mains power quality should be that of a typical
for 0,5 cycles
commercial or hospital environment.
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
If the user of the device requires continued operation
(30 % dip in UT)
during power mains interruptions, it is recommended
for 25 cycles
that the device be powered from an uninterruptible
power supply or a battery.
< 5 % UT
(> 95 % dip in UT)
for 5 seconds
Magnetic power frequency fields should be that of a
3 A/m
typical commercial or hospital environment.
Electromagnetic environment guide