ELECTROMAGNETIC COMPATIBILITY
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments
Warning: Don't near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance
imaging, where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
Warning Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and
result in improper operation."
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the equipment, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Technical description:
1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic
disturbances for the excepted service life.
2. Guidance and manufacturer's declaration -electromagnetic emissions and Immunity.
Guidance and manufacture's declaration - electromagnetic emissions
Emissions Test
RF emissions CISPR11
RF emissions CISPR11
Harmonic distortion
IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
32 • ENGLISH
Table 1
Compliance
Group 1
Class B
Class A
Compily