CARE & SAFETY INFORMATION
Operation outside of stated operating temperature may result in measurement
error or device malfunction; operation environment temperature is: 41°F – 104°F
(5°C – 40°C); Humidity: 15% to 90% RH (non-condensing).
Never use this unit while operating a vehicle or in the bathtub or shower.
To avoid measurement errors, please avoid the condition of strong
electromagnetic field radiated interference signal or electrical fast transient/burst
signal.
The blood pressure monitor, its adaptor, and the cuff are suitable for use within
the patient environment. If you are allergic to polyester, nylon or plastic, please
don't use this device.
During use, the patient will be in contact with the cuff. The materials of the cuff
have been tested and found to comply with requirements of ISO 10993-5:2009
and ISO 10993-10:2010. This cuff will not cause any potential sensation or
irritation reaction.
Wireless communications equipment such as wireless home network devices,
mobile phones, cordless telephones and their base stations, walkie- talkies
can affect this equipment and should be kept at least a distance d away from
the equipment. The distanced is calculated by the MANUFACTURER from the
80MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as
appropriate.
The equipment is not AP/APG equipment and not suitable for use in the presence
of a flammable anesthetic mixture with air of with oxygen or nitrous oxide.
On the rare occasion of a fault causing the cuff to remain fully inflated during
measurement, open the cuff immediately. Prolonged high pressure (cuff pressure
>300mmHg or constant pressure >15mmHg for more than 3 minutes) applied to
the arm may lead to an ecchymosis.
If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically
deflate. Should the cuff not deflate when pressures reaches 40 kPa (300 mmHg),
detach the cuff from the arm and press any button to stop inflation.
The operator shall not touch output of batteries /adapter and the patient
simultaneously.
6 • ENGLISH