Memory Functions; Recalling Measurements In Memory - smartheart 01-509 Manual De Instrucciones

Ocultar thumbs Ver también para 01-509:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MEMORY FUNCTIONS

Recalling Measurements in Memory

The switch at the top of the device must be in
the ON position to review the memory banks; it
will not allow review if not in the ON position.
1.
When the monitor is off, press and release
the MEMORY
memory bank.
Select U1/SET or U2 to review the
2.
respective readings in memory.
3.
The unit will first display the average of
the last three measurements, Fig 1. If there
are less than 3 readings in memory, the
most recent reading will display.
Press the MEMORY
4.
begin scrolling through the saved
measurements. Measurements will appear
on the display from most current to oldest;
the memory number alternate in the same
position as the date/time stamp.
5.
All results for a given measurement will
display, including measurement results,
pulse rate, Hypertension Indicator and
Irregular Heartbeat alert (if applicable).
The date/time stamp of the measurement
will alternate as well, Fig 2.
6.
When the number of readings exceeds
250, the oldest data will be replaced with
the new record.