Integra MAYFIELD A2010 Manual De Instrucciones página 7

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 60
The MAYFIELD Adult Radiolucent Horseshoe Headrest is designed to be used with the included Horseshoe Gel Pads. Replacement Gel
Pads are available individually (REF 41C1444 Adult Left and 41C1445 Adult Right).
The included Pulley Rod can be attached for traction of up to 40 lbs. if desired, or left unattached.
Use of MAYFIELD products and accessories in conjunction with other manufacturers' stabilization equipment is not recommended.
Inspection
Always inspect instruments before and after use. If a component appears damaged and/or does not seem to function properly, do not use
the device and immediately send the instrument to an authorized Integra repair center for evaluation, repair or replacement. Allow your
Integra Representative to inspect this device a minimum of two times per year to assist you in its proper function.
WARNING:
Device must not be used if it appears to be damaged or functioning incorrectly. Doing so may result in serious patient injury.
Gel Pads
1. Pads should not have any rips or tears in the top surface.
2. Backs of the pads should not be broken or cracked.
Stationary Horseshoe Faceplate with attached Pulley Rod Knob
1. Pulley Rod Knob should freely turn and easily screw into the threaded hole of the Pulley Rod.
2. Small starburst teeth of the Stationary Horseshoe Faceplate should match up and tighten securely with the starburst teeth of the
Pulley Rod.
3. Large starburst teeth of the Stationary Horseshoe Faceplate should match up and tighten securely with the starburst teeth of the
Radiolucent Short Swivel Adaptor or Radiolucent Tri-Star Swivel Adaptor.
4. Parts should be smooth and free of cracks and voids.
5. Gel Pad should slide easily onto the Stationary Horseshoe Faceplate and snap in place.
Slide Horseshoe Faceplate with attached Adjustment Knob
1. Slide Horseshoe Faceplate should slide easily onto the Stationary Horseshoe Faceplate.
2. Adjustment Knob should turn easily and securely lock down on the slide bar of the Stationary Horseshoe Faceplate.
3. Parts should be smooth and free of cracks and voids.
4. Gel Pad should slide easily onto the Slide Horseshoe Faceplate and snap into place.
Pulley Rod
1. Part should be smooth and free of cracks and voids.
Instructions for Use
Mounting of Base Unit and attachment of Transitional Members and Swivel Adaptor
IMPORTANT:
Refer to the corresponding MAYFIELD instruction manual (REF 451A2004, 451A2079) for the correct usage of the applicable
MAYFIELD Base Unit prior to using this device. If applicable, refer to the corresponding MAYFIELD Tri-Star Swivel Adaptor
instruction manual (REF 451A1111, 451A2111) prior to using this device.
1.
Mount the MAYFIELD Radiolucent Base Unit to the operating table.
2. Attach the Transitional Members and the Short Swivel Adaptor to the Radiolucent Base Unit. If extra starburst attachments for
ancillary equipment are desired, the MAYFIELD Radiolucent Tri-Star Swivel Adaptor may be used instead of the Short Swivel Adaptor.
Figure 3
6
Figure 4
EN – ENGLISH

Publicidad

Tabla de contenido
loading

Tabla de contenido