EN – ENGLISH
Indications for Use
The MAYFIELD® Adult Radiolucent Horseshoe Headrest is intended to be used to support a patient's head in both the prone and supine
positions during diagnostic examination or surgical procedures where support and stabilization of the head is desired but complete
fixation of the head is not required.
WARNING:
Failure to read and follow instructions furnished in this product insert may result in serious patient injury..
WARNING:
Device must not be used if it appears to be damaged or functioning incorrectly. Doing so may result in serious patient injury.
WARNING:
Ensure the mating starburst teeth of the Swivel Adaptor and the Stationary Horseshoe Faceplate are the same size and have
properly engaged (meshed) when fully tightened. Failure to do so may damage teeth, compromise stability, and result in serious
patient injury.
WARNING:
Do not position the Slide Horseshoe Faceplate past the end of the Stationary Horseshoe Faceplate slide bar (Figure 11). Doing so
may result in serious patient injury.
WARNING:
Horseshoe Faceplates should be adjusted prior to patient positioning. Make certain to always support the patient's head when
adjusting the Horseshoe Faceplates or unlocking any connection of this device. Failure to do so may result in serious patient
injury.
WARNING:
Failure to properly position the patient and to fully secure all adjustment portions of this or any support device may result in
serious patient injury.
WARNING:
Pulley Rod may be used for traction of up to 40 lbs. Exceeding 40 lbs. of traction load may result in device failure and serious
patient injury.
WARNING:
Ensure the mating starburst teeth of the Pulley Rod and the Stationary Horseshoe Faceplate are the same size and have properly
engaged (meshed) when fully tightened. Failure to do so may damage teeth, compromise stability and support of traction load,
and result in serious patient injury.
WARNING:
Device must not be used if it has been dropped or mishandled. Doing so may result in serious patient injury.
WARNING:
Do not steam sterilize any component of this device. Doing so may damage device and result in serious patient injury.
4
EN – ENGLISH