Power off suddenly when inflating
Hold the on/off button but can not start
the device
Cuff inflation start before press the measurement button or never stop inflating
when measuring
Cuff never deflation
Air pressure error
No press value displayed or the value unchanged or change erratically when
cuff inflated
Other phenomenon
No use for a long time, the power of
batteries can be exhausted owing to
the changed temperature
Power of batteries can be exhausted
The battery polarities is reversed
No deflation or deflation error or
inflation without stop
Others
ENGLISH
Replace all four batteries with new
ones.
Replace all four batteries with new
ones.
Check the battery installation for prop-
er placement of the battery polarities.
Pull out the cuff to deflate.
Stop using the device and contact us.
Pull out the cuff to deflate.
Stop using the device and contact us.
Pull out the cuff to deflate.
Stop using the device and contact us.
Keep arm, body still, measure again
Pull out the cuff to deflate.
Stop using the device and contact us.
Switch on the power once again and
restart an operation.
Replace the batteries.
If no, please contact us.