Mechanical/Automated Cleaning

Detergents
• A neutral pH enzymatic detergent is recommended.
• Alkaline detergents with a maximum pH of 11 may be used but user must
demonstrate their effectiveness.
• The process, detergent and any additives must be discussed with detergent
manufacturer.
Caution: Do not use any alkaline detergent with a pH greater than 11. Alkaline
detergents attack grease and seals which can increase wear and cause device
to malfunction.
Note: Use of alkaline detergents can cause color to fade but this does not impair
function of device.
High Speed Handpieces and Attachments
1. Loading the Basket
• Load devices in basket by securing them in appropriate brackets as marked
on the basket with distal end angled down.
Warning: Red protective cap for Perforator Driver (CSR60) must be left off for
mechanical/automated cleaning.
Caution: Visually inspect electrical contacts and cable of electric handpieces for
damage. Connector ends of electric & pneumatic handpieces must be sealed
with special seal caps provided. See Figure 6.
Note: Before putting on seal cap, turn black cap to wrap cable tight around it.
Figure 6
2. Mechanical/Automated Cleaning Cycle Process Parameters
Cycle
Rinse
Pre-wash
Cleaning
Rinse
Thermal Processing 5 minutes
Dry
Caution: Do not use neutralizing agent.
3. Inspect
• Visually inspect devices. Repeat manual pre-clean and mechanical/
automated clean if visible soil remains.
• Remove seal cap from connector end of handpiece and drain until no visible
droplets are coming from the handpiece or connector.
4. Lubricate
Caution: This step is for attachments only. Do not lubricate handpieces.
Option 1: 05.001.078, Lubricant for Anspach Systems, 110 ml
Warning: Do not use Gravity Air Displacement Steam Autoclave Sterilization
Process.
8
Duration (minimum) Cleaning Instructions
2 minutes
1 minute
2 minutes
5 minutes
40 minutes
Cold deionized (DI) or purified water (PURW)
Warm DI or PURW water (≥ 40°C); use detergent
Warm DI or PURW water (≥ 45°C); use detergent
Rinse with DI or PURW water
Hot DI water, ≥ 93°C
≥ 90°C