• For the following attachments apply one (1) spray pump spray to the
proximal end of the attachment.
SHORT
SHORT-HD
MEDIUM
MEDIUM-HD
• For the following attachments apply one(1) spray pump spray to the
distal end of the attachment and one (1) to the proximal end of the
attachment.
MA-D20
CSR60
• Applying excess amounts of lubricant will cause the lubricant drip from
the attachment. Clean off any excess lubricant.
Option 2: Non-silicone based medical lubricant
• Prepare a lubricating solution of instrument milk (non-silicone based
medical lubricant) per the manufacturer's directions.
• Fully immerse attachment in lubricating solution, at room temperature,
in a suitable container and agitate for 15 seconds.
Caution: Do not rinse out instrument milk (lubricant). Do not apply
mineral oil or other lubricants, which may cause attachment to overheat.
5. Go to Sterilization section for further instructions.
Note: The clinical processing instructions provided have been validated by
DePuy Synthes Power Tools for preparing a nonsterile medical device; this
instruction is provided in accordance with ISO 17664 and ANSI/AAMI ST81.

Sterilization

Before sterilizing: Ensure sterilization equipment is in proper working
order as specified by manufacturer. Assure equipment manufacturer's
instructions are properly employed by trained and qualified personnel.
Assure actual cycle employed has been properly validated for the
device(s)/load configuration being processed and appropriate sterilization
indicator devices are included for each process and cycle.
1. Loading the Basket
• Load attachments in EM/XM Cleaning and Sterilization Basket by
securing them in appropriate brackets as marked on basket with distal
end angled down.
• Replace seal cap in its bracket and lay handpiece connector in bottom
of basket.
Warning: Red protective cap for CSR60 must be left off for sterilization.
Seal caps must be removed from connector ends of electric and pneumatic
handpieces prior to sterilization. Devices must be cleaned prior to sterilization
using the validated cleaning method outlined in these instructions.
Caution: Black cap must remain on connector end of handpiece.
2. Packaging for Sterilization
• When wrapping the basket the lid must be used.
• Double wrap basket or instruments in accordance with local
procedures, using standard wrap and wrapping techniques such as
those described in ANSI/AAMI ST79.
Warning: When using a sterilization wrap the basket lid must be used. You
must follow manufacturer's instructions for any wraps used. Do not use
individual sterilization pouches. Only use indicated basket for equipment.
LONG
CRANI-L
LONG-HD
CRANI-P
MIA16
MA-10S
CRANI-A
MA-10C
QD8
QD11
MA-15S
MA-15C
MA-15ST
MA-19ST