Descargar Imprimir esta página

Orthofix TrueLok 54-1139 Manual Del Usuario página 2

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 10
PQ WTN D 10/18
EN
1) The Wire Tensioner (54-1139) allows tensioning wires up to 130kg. An indication of the applied tension can be read from a scale (fig. 1).
2) Only when tensioning 1.5mm diameter wires, apply the specific Tension Limiter (54-1141) to physically limit the applied tension to 90kg.
Make sure to insert the Tension Limiter correctly, paying attention to the laser mark (fig. 2).
3) To exchange the different types of Wire Tensioner Tips, remove the Wire Tensioner's Tip by loosening the locking set-screw with the hex-driver (52-1018) (fig. 3). Do not remove the locking set-screw for any reason.
4) Using appropriate tools thoroughly clean the whole cannulation (fig. 4) to remove any debris that might be present. Additional information on cleaning and reprocessing can be found in the leaflet PQTLK.
5) Apply one drop of medical grade lubricant on moving elements when needed.
NOTE: Products that show excessive fading of marked product code and lot, thus preventing clear identification and traceability, must NOT BE USED.
Instructions For Use are subject to change; the most current version of each Instruction For Use is always available online.
Important information - please read prior to use
Keep this leaflet: you may need to read it again
TrueLok WIRE TENSIONER (54-1139)
Figure 4
Manufacturer name
ORTHOFIX SRL, via Delle Nazioni 9, 37012 Bussolengo (VR) Italy
Tel. 0039 (0) 45 6719000 - Fax 0039 (0) 45 6719380
Figure 1
Figure 2
Figure 3
2