Automatic Cleaning; Inspection/Maintenance - CareFusion Snowden-Pencer Guia De Inicio Rapido

Ocultar thumbs Ver también para Snowden-Pencer:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. For lumen devices, place the device into the open/relaxed position with the distal tip
pointed down. Flush the device with a minimum of 50 ml of detergent solution utilizing
a temperature range of 27°C to 44°C (81°F to 111°F), by using the flushing port
located on the handle/shaft. Repeat the flush process a minimum of 2 times (i.e. total
of 3 times); ensuring all fluid exiting the lumen is clear of soil.
7. For lumen devices, if visible soil is detected during the final lumen flush, re-perform
brushing and flushing of the lumen.
Note: The final rinse steps (8 and 9) should be performed with treated water:
deionized, distilled or reverse osmosis.
8. Rinse the device by completely immersing it in treated water with a temperature range
of 27°C to 44°C (81°F to 111°F), for a minimum of 30 seconds to remove any residual
detergent or debris.
9. For lumen devices, following the rinsing step, place the device into the open/relaxed
position with the distal tip pointed down. Flush the device with a minimum of 50 ml of
treated water utilizing a temperature range of 27°C to 44°C (81°F to 111°F), by using
the flushing port located on the handle/shaft. Repeat the flush process a minimum of 2
times (i.e. total of 3 times).
10. Dry the device with a clean, lint-free towel.
11. For lumen devices, manipulate the device to allow rinse water to drain from the lumen.
12. Visually examine each device for cleanliness.
13. If visible soil remains, repeat cleaning procedure until the device is thoroughly clean.

Automatic Cleaning

1. Ensure all pre-processing instructions are followed prior to cleaning.
2. Clean the devices via the Automatic cleaning parameters below.
PHASE
RECIRCULATION TIME
Pre-Wash
Enzyme Wash
Wash
Rinse
Pure Rinse
Drying
3. For lumen devices, manipulate the device to allow rinse water to drain from the lumen.
4. If visible moisture is present, dry the instrument with a clean, lint-free towel.
5. Visually examine each instrument for cleanliness.
6. If visible soil remains, repeat cleaning procedure until the device is thoroughly clean.

Inspection/Maintenance

Proper care and handling is essential for satisfactory performance of any surgical device.
The previous cautions should be taken to ensure long and trouble-free service from all
your surgical devices. Inspect devices before each use for broken, cracked, tarnished
surfaces, movement of hinges, and chipped or worn parts. If any of these conditions
appear, do not use the device. Return devices to an authorized repair service center for
repair or replacement.
Before sterilizing, lubricate the device with instrument milk or a steam permeable/ water
soluble lubricant, following the lubricant manufacturer's instructions. Check for smooth
action of all moving parts.
Immediately proceed to recommended steam sterilization. Do not store instrument prior
to sterilization.
Let devices drip dry for three (3) minutes before packaging for sterilization.
Packaging
Devices can be loaded into dedicated packaging systems. Sterilization wrap material must
be cleared for the applicable sterilization modality by your countries regulatory body.
Use in accordance with packaging manufacturer's sterilization instructions being sure to
protect jaws and cutting edges from damage.
MINIMUM
15 Seconds
1 Minute
2 Minutes
15 Seconds
10 Seconds
N/A
WATER
TEMPERATURE
Cold Drinking/Potable
Water
1°C - 16°C
(33°F - 60°F)
Hot Drinking/Potable
Water
43°C - 82°C
(110°F - 179°F)
Hot Drinking/Potable
Water
43°C - 82°C
(110°F - 179°F)
Hot Drinking/Potable
Water
43°C - 82°C
(110°F - 179°F)
Purified Water (Hot)
43°C - 82°C
(110°F - 179°F)
N/A
3
DETERGENT TYPE AND
CONCENTRATION
(IF APPLICABLE)
N/A
• Detergent: pH-neutral/
enzymatic
• Concentration: Per the
detergent manufacturer's
recommendations
• Detergent: pH-neutral
cleanser
• Concentration: Per the
detergent manufacturer's
recommendations
N/A
N/A
N/A

Publicidad

Tabla de contenido
loading

Tabla de contenido