Phase
Rinse 1
Thermal
Rinse**
** Optional phase for disinfection of components – minimum water
temperature as indicated or per worker manufacturer specifications
for the thermal rinse cycle.
NOTE – Any deviation from this guideline could result in damage to
the equipment as well as improper cleaning results.
Rinse with purified water. Do not perform if parameters cannot be
achieved.
1. Remove from washer and dry completely if needed.
2. Inspect equipment to make sure there is no visible organic debris
or residue from cleaning agent.
Repeat process if any visible soil is detected.
NOTE: These pins are supplied non-sterile and must be sterilized
prior to each use.
Sterilization
Inspect pins carefully under a strong light source. Pin points must
be sharp and straight. O-rings must be smooth and free of tears.
Remove damaged pins from service.
Cycle
Preconditioning Pulses:
Temperature:
Duration:
Dry time:
Other time and steam temperature cycles may also be used; however,
the user must validate any deviation from the recommended time
and temperature. Follow the sterilizer manufacturer's instruction for
operation and loading of the steam autoclave.
Time
Water Tem-
(Min.)
perature
0:30
Hot water
82.2ºC
2:00
(180º F)
Detergent and
Concentration
N/A
N/A
Pre-vacuum
3
132°C (min)
3 minutes
5 minutes
10